Calibration Kit H28A
UCP1A/UCP1B Probe Calibration Kit
Description
The Buckleys H28A (6004-0032) UCP1A/UCP1B Probe Calibration Kit is a calibration kit for verifying the accuracy of the readings on Buckleys Contact and Proximity Probes, so that in-house verification certificates can be issued. The kit contains the components and full instructions for this in-house verification.
When the probe readings are correct, the silver/silver chloride screw-in half cell is checked against K-Series Reference Cells in a standard salt water solution. To confirm the readings, three cells are used and the sea water corrosion test mixture is made up to a 3 % solution. The technical specifications state an output of +208 mV @ 25 °C vs SHE, an accuracy of ±5 mV @ 25 °C vs SHE, a temperature coefficient of −0.64 mV per °C and a temperature range of 0–40 °C.
The Buckleys H28A kit comprises 3 screw-on K-Series reference electrode cells, sea water corrosion test mixture, 3 adaptor leads, 1 zinc block, 1 magnesium block and full instructions for in-house verification. All kit parts can be replaced individually. The packed calibration kit weighs 1.2 kg and measures 26 × 14 × 30 cm.